    Jian Chen, Hao Wu, An-wen Liu, Shui-ming Hu, Jingsong Zhang. Field Measurement of NO2 and RNO2 by Two-Channel Thermal Dissociation Cavity Ring Down Spectrometer[J]. Chinese Journal of Chemical Physics , 2017, 30(5): 493-498. DOI: 10.1063/1674-0068/30/cjcp1705084
    Citation: Jian Chen, Hao Wu, An-wen Liu, Shui-ming Hu, Jingsong Zhang. Field Measurement of NO2 and RNO2 by Two-Channel Thermal Dissociation Cavity Ring Down Spectrometer[J]. Chinese Journal of Chemical Physics , 2017, 30(5): 493-498. DOI: 10.1063/1674-0068/30/cjcp1705084

    Field Measurement of NO2 and RNO2 by Two-Channel Thermal Dissociation Cavity Ring Down Spectrometer

    • A two-channel thermal dissociation cavity ring down spectroscopy (CRDS) instrument has been built for in situ, real-time measurement of NO2 and total RNO2 (peroxy nitrates and alkyl nitrates) in ambient air, with a NO2 detection limit of 0.10 ppbv at 1 s. A 6-day long measurement was conducted at urban site of Hefei by using the CRDS instrument with a time resolution of 3 s. A commercial molybdenum converted chemiluminescence (Mo-CL) instrument was also used for comparison. The average RNO2 concentration in the 6 days was measured to be 1.94 ppbv. The Mo-CL instrument overestimated the NO2 concentration by a bias of +1.69 ppbv in average, for the reason that it cannot distinguish RNO2 from NO2. The relative bias could be over 100% during the afternoon hours when NO2 was low but RNO2 was high.